Primary objective: To compare the number of episodes of atrial fibrillation detected by the Corsano CardioWatch 287-2 during 28 days of use with the number of episodes of atrial fibrillation detected by conventional Holter monitoring during 48 hours…
ID
Source
Brief title
Condition
- Cardiac arrhythmias
- Vascular hypertensive disorders
Synonym
Research involving
Sponsors and support
Intervention
Outcome measures
Primary outcome
Absolute percentage increase of patients in whom at least one event of atrial
fibrillation is detected by the Corsano CardioWatch 287-2 during 28 days of
monitoring, compared to a conventional EKG holter during 24-48 hours of
monitoring.
Secondary outcome
(1) Mean blood pressure difference and its SD between blood pressure measured
by the Corsano CardioWatch 287-2 and blood pressure measured by a conventional
oscillometric blood pressure cuff.
(2) Usability of the Corsano CardioWatch 287-2 in a remote care setting as
determined by a questionnaire.
(3) Difference between awake-asleep blood pressure change as measured by the
Corsano CardioWatch 287-2 and awake-asleep blood pressure change as measured by
a conventional 30-min. oscillometric blood pressure cuff across 24 hours.
Background summary
Wearables have the potential to monitor patients remotely. The Corsano
CardioWatch 287-2 is such a medical device that can monitor atrial fibrillation
and long-term blood pressure. The device has been validated using clinical
trials in hospitals, but validation in the intended remote setting is lacking.
Study objective
Primary objective: To compare the number of episodes of atrial fibrillation
detected by the Corsano CardioWatch 287-2 during 28 days of use with the number
of episodes of atrial fibrillation detected by conventional Holter monitoring
during 48 hours of use (standard care).
Secondary objective: To assess the difference in blood pressure measurements
obtained by the Corsano CardioWatch 287-2 and the conventional cuff blood
pressure monitor over a period of 28 days; to assess the usability of the
Corsano CardioWatch 287-2 from a patient perspective; to assess the difference
between awake-asleep blood pressure change as measured by the Corsano
CardioWatch 287-2 and awake-asleep blood pressure change as measured by a
conventional 30-min. oscillometric blood pressure cuff across 24 hours.
Study design
Group A:
Conventional 24 hour EKG holter monitoring will be compared to 28 day
continuous monitoring by the Corsano CardioWatch 287-2 to assess whether the
percentage of patients in whom atrial fibrillation is detected will increase.
Detection of atrial fibrillation with the Corsano CardioWatch 287-2 is done by
optical photoplethysmography (PPG), after which atrial fibrillation is
confirmed with a 30-second ECG measurement.
Additionally, conventional cuff blood pressure will be compared to blood
pressure measured by the Corsano CardioWatch 287-2 to validate free-living
blood pressure monitoring. Patients will measure their blood pressure daily
with a cuff blood pressure measurement device. The measurements at day 7, 14,
21 will be used to reinforce the Corsano blood pressure algorithm. The
remaining measurements will be used to compare the blood pressure with the
values measured by the Corsano CardioWatch 287-2 to assess accuracy.
Group B:
Conventional automatic cuff blood pressure measurement will be compared with
the Corsano CardioWatch 287-2 for 24-48 hours. The automatic cuff blood
pressure monitor will take a blood pressure reading every 30 minutes. Each
reading will be compared with the Corsano CardioWatch 287-2 reading.
Study burden and risks
At the cardiologist's office, Group A and B patients receive 3 blood pressure
measurements. These are performed by a nurse. Group A will take daily blood
pressure readings of themselves at home. If atrial fibrillation is detected by
the wristband, Group A patients will be asked to take an ECG. Finally, Group A
patients will be asked once to complete a questionnaire. Group B only has to
wear the wristband for 24 to 48 hours as an additional burden in addition to
the aforementioned initial measurements.
Wilhelmina van Pruisenweg 35
Den Haag 2595 AN
NL
Wilhelmina van Pruisenweg 35
Den Haag 2595 AN
NL
Listed location countries
Age
Inclusion criteria
* >= 18 years old;
* able to provide consent;
* receiving EKG holter or automatic blood pressure cuff for home monitoring per
doctor prescription
Exclusion criteria
* unable to wear the Corsano CardioWatch 287 due to reasons such as allergic
reactions, wounds, amputations etc.;
* unable to receive blood pressure measurements per cuff due to lymphedema,
amputation, dyalisis shunt, wounds, etc.;
* pregnant women;
* breastfeeding women;
* upper arm circumference not within the cuff range (22-42 cm)
* unable or not willing to sign informed consent;
* significant mental or cognitive impairment
Design
Recruitment
Medical products/devices used
metc-ldd@lumc.nl
metc-ldd@lumc.nl
Followed up by the following (possibly more current) registration
No registrations found.
Other (possibly less up-to-date) registrations in this register
No registrations found.
In other registers
Register | ID |
---|---|
CCMO | NL83281.000.22 |